The GPR-7500 analyzer utilizes the low-power electro-chemical sensor technology to provide a cost-effective and easy to maintain online-installation hydrogen sulfide measurement in a small package. The GPR-7100 is a portable H2S analyzer version preferably used for spot checking the process or temporary usage during servicing or commissioning of a permanent installation gas analyzer. These advanced instruments from AII are agency approved for use in flammable gas streams.
AII GPR-7500 PPM H2S Analyzer
-Accuracy: 2% of FS range under constant conditions
Analysis: 0-20, 0-50, 0-100 PPM FS ranges; Auto-ranging or manually lock on single range
Application: Trace H2S in flowing gas streams
Area Classification: Certified for use in hazardous areas - see lower right
Enclosure: NEMA Type 3R for rain in outdoor applications (UL) NEMA 4X (ATEX)
Flow: Maintain constant flow of Sample and Air at 1 SCFH; maintain Sample to Air ratio of 1:1
Pressure: Inlet - regulate to 5-30 psig to deliver constant 1 SCFH flow; vent – atmospheric
Power: 12-28 VDC Transmitter, 12 VDC Pump; 12-28 VDC (UL or ATEX Certified) 110-220 VAC (ATEX Certified) Transmitter, 12-24 VDC PUMP
Sample System: Coalescing filter, flow meters, sample/span valve; Flow through SS sensor housing, flow control valve, sample flow meter, 3-way sample/span valve, air pump, air flow meter
Sensor Model: OSV-72-7HH (standard ranges)/ OSV-72-7HH-LD (standard ranges)
Temp. Range: -10 ºC to 45ºC