The GPR-1800 series of intrinsically safe trace oxygen analyzers are designed for use in demanding process environments where the low detection limit down the AII’s industrial oxygen sensors can be well utilized. Typical for Analytical Industries gas analyzers, they are simple to use as well as easy and cost-effective to maintain thanks to the use of their maintenance-free galvanic oxygen sensors. The innovative liquid drain manifold, which is available as an option, protects and extends the sensor's life span in processes where liquids may be present in the sample gas.
AII GPR-1800 PPM Oxygen Analyzer
-Accuracy: < 2% of FS range under constant conditions
Analysis: 0-10, 0-100, 0-1000 PPM, 0-1%, 0-25% (CAL) FS Auto-ranging or manual lock on a single range
Application: Oxygen analysis in inert, hydrocarbon, helium, hydrogen, mixed and acid (CO2) gas streams
Area Classification: Certified for use in hazardous areas
Connections: 1/8" compression tube fittings
Enclosure: NEMA Type 3R suitable for rain in outdoor applications (UL); NEMA 4X (ATEX)
Flow: Not flow sensitive; recommended flow rate 1-2 SCFH
Pressure: Inlet - regulate to 5-30 psig to deliver 1-2 SCFH flow; vent - atmospheric
Power: 12-28 VDC (ULc and ATEX Certified); 110-220 VAC (ATEX Certified); 18-24 VDC
Sample System: Sample flow meter; Unique liquid drain sensor manifold, flow indicator; options available, see other side
Sensor Model: GPR-12-333-LD for non-acid (CO2) gas streams; XLT-12-333-LD for gases containing > 0.5% CO2; XLT-12-333 for gases containing > 0.5% CO2
Operating Range: 5ºC to 45ºC (GPR sensor); -10ºC to 45ºC (GPR sensor); -10º to 45ºC (XLT sensor)